Herunterladen Diese Seite drucken

TOPRO TAURUS E Bedienungsanleitung Seite 14

Vorschau ausblenden Andere Handbücher für TAURUS E:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Guidance and manufacturer's declaration
Electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS
The TOPRO Taurus is intended for use in the electromagnetic environment specified below.
The customer or the user of the TOPRO Taurus should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to earth
Voltage dips, short
<5 % UT
interruptions and
(>95 % dip in UT)
voltage variations
for 0,5 cycle
on power supply
40 % UT
input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration
Electromagnetic IMMUNITY for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
The TOPRO Taurus is intended for use in the electromagnetic environment specified below.
The customer or the user of the TOPRO Taurus should assure that it is used in such an environment.
IMMUNITY
IEC 60601
test
TEST LEVEL
Conducted
3 Vrms
RF
150 kHz to 80
IEC 61000-
MHz
4-6
12
Compliance level Electromagnetic environment – guidance
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be
used no closer to any part of the TOPRO Taurus, including
cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d=1,16¥P
3 V
d=1,16¥P 80 MHz – 800 MHz
d=2,33¥P 800 MHz – 2,5 GHz
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a
typical commercial or hospital environment.
If the user of the TOPRO Taurus requires
continued operation during power mains
interruptions, it is recommended that the
TOPRO Taurus be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.