Laboratory pump Laboport® N96
Translation of Original Operating Instruction, KNF 323241-323242 06/22
g.) Final steps
WARNING
1. If necessary, screw in attachment parts such as silencer,
suction filter, etc., from the gas connections.
2. Perform a function test.
à
Connect the gas connections to the pump.
à
Connect the pump to the power supply.
à
Test the pump for proper function (among other
things, ultimate vacuum).
à
Again disconnect the pump electrically and pneumati-
cally.
3. Install the pump in the desired installation location accord-
ing to 7 Commissioning [} 19].
à
Connect the gas connections to the pump.
à
Connect the pump to the power supply.
à
Test the pump for proper function (among other
things, ultimate vacuum).
If you have questions with regard to maintenance, please
contact your KNF Customer Service (contact data: see
www.knf.com).
Risk of injury and poisoning from leaks
à
Before recommissioning the
pump, check the pump heads and
pneumatic connections for leaks.
Leaks may lead to poisoning,
chemical burns or similar injuries.