Drying
• Carefully dry all components using a soft cloth.
• Shake the mesh atomiser (3) lightly backwards and
forwards (5 to 10 times), so that the water inside
the mesh is removed from the small holes.
• Place the individual components on a dry, clean
and absorbent surface and allow to dry completely
(for at least 4 hours).
Note
Ensure that all parts are thoroughly dried after cleaning,
otherwise there is an increased risk of bacterial growth.
Once all components are fully dry, reassemble the
nebuliser and place the components in a dry, sealed
container.
Ensure that the mesh atomiser (3) has dried completely
by shaking it. Otherwise, nebulisation may not function
properly after reassembling the device.
If this is the case, disassemble the mesh atomiser (3)
again and shake it so that the water can escape. Once
it has been reassembled, the nebuliser should function
properly again as usual.

Material resistance

• When choosing a cleaning or disinfecting agent,
note the following: Only use mild cleaning or disin-
fectant agents in the quantities recommended by
the manufacturer.
• As with all plastic components, frequent use and
cleaning of the atomiser and the accessories can
lead to a certain amount of wear. Over time, this
can alter the aerosol properties and may eventually
affect the efficiency of therapy. We therefore rec-
ommend that you replace the atomiser and acces-
sories at least once a year.
Storage
• Do not store the nebuliser in a damp atmosphere
(e.g. in the bathroom) or transport it together with
damp objects.
• Store and transport the nebuliser away from direct
sunlight.
Troubleshooting
Problem/
Possible Cause/Remedy
Question
The nebuliser
1. Insufficient medicine in the
produces little
atomiser.
or no aerosol.
2. The atomiser is not held in an up-
right position.
3. The medicine used is unsuitable
for spraying (e.g. too thick, viscos-
ity must be less than 3).
The medicine solution should be
specified by the doctor.
Problem/
Possible Cause/Remedy
Question
Output is too
1. The batteries are flat. Replace the
low.
2. There are air bubbles inside the
3. Particles on the mesh are imped-
4. The mesh is worn out.
Which medi-
Only the doctor can advise you
cines are suit-
which medicine to use to treat your
able for inhala-
condition. Consult your doctor.
tion?
With the JIH 50, you can atomise
medicines with a viscosity lower
than 3.
Some inhala-
This is normal and occurs for techni-
tion solution
cal reasons. Stop inhaling as soon
remains in the
as you can hear a marked differ-
nebuliser.
ence in the sound made by the at-
omiser, or when the device switches
off automatically due to insufficient
inhalant.
What special
1. In babies, the mask should cover
steps should
be taken for
babies and
2. In children, the mask should also
children?
Note: Children should only use the
device with help and under supervi-
sion of an adult. Never leave a child
alone with the nebuliser.
Inhalation with
This is for technical reasons. Less
the mask takes
medicine is inhaled per breath
longer.
through the mask holes than us-
ing the mouthpiece. The aerosol is
mixed with ambient air through the
holes in the mask.
Does each user
This is absolutely necessary for hy-
need their own
giene reasons.
accessories?
18
batteries or connect the power
supply unit and try again.
medicine container which are pre-
venting the medicine from coming
into continuous contact with the
mesh. Please check and remove
any air bubbles.
ing output. To remove the par-
ticles, fill the nebuliser with 2 to
3 drops of vinegar and 3 to 6 ml
of water, then nebulise this solu-
tion completely. Do not inhale this
spray and ensure that you clean
and disinfect the medicine con-
tainer afterwards (see page 16).
If the output does not improve, re-
place the mesh.
the mouth and nose to guarantee
effective inhalation.
cover the mouth and nose. Nebu-
lisation next to a sleeping person
is not suitable because insufficient
medicine reaches the lungs.