Attaching the Replacement Hose
1.
Attach the cable connector of the new hose to the header plug, making sure the protrusions on the
connector align with the corresponding slots on the header.
2. Align the flat section of the hose end with the corresponding flat part of the unit's hose connector and slide
the replacement hose into the temperature management unit.
3. Firmly screw the hose collar back into place.
4. Reconnect the temperature management unit to a grounded power source.
5. Calibrate operating temperature settings.
6. Perform over-temperature detection system test.
NOTE: See the Bair Hugger Model 775 Preventive Maintenance Manual for all service and
testing procedures.
Symbol Glossary
Symbol Title
Manufacturer
Date of
manufacture
Batch code
Catalogue number
Consult
instructions for
use or consult
electronic
instructions for
use
For more information see, HCBGregulatory.3M.com
Symbol
Description and Reference
Indicates the medical device manufacturer. Source: ISO 15223, 5.1.1
Indicates the date when the medical device was manufactured.
Source: ISO 15223, 5.1.3
Indicates the manufacturer's batch code so that the batch or lot can be
identified. Source : ISO 15223, 5.1.5
Indicates the manufacturer's catalogue number so that the medical device can
be identified. Source : ISO 15223, 5.1.6
Indicates the need for the user to consult the instructions for use.
Source: ISO 15223, 5.4.3