Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications - Omron HEM-907 Gebrauchsanweisung

Digitales automatisches blutdruckmessgerät
Vorschau ausblenden Andere Handbücher für HEM-907:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 141

Specifications

Product category
Product description
Model (code)
Display
Measurement
Measurement Range
Accuracy
Inflation
Deflation
Air Release
Rating
Power supply
Protection Class
Degree of Protection
Operating conditions
Storage/Transport conditions
Weight of Main Unit
External Dimensions
Also included in the package
Options
Please understand that specifications may be changed without prior notice.
This device is clinically investigated according to the requirement of ESH IP:2002.
This device can be used for continuous operation.
This blood pressure monitor conforms to the European standard EN 1060, Non-invasive Sphygmomanometers. Part 1:
General Requirements and Part 3: Additional Requirements for Electromechanical Blood Pressure Measuring Systems.
Please report to the manufacturer and the competent authority of the Member State in which you are established about
any serious incident that has occurred in relation to this device.
24
:
Electronic Sphygmomanometers
:
OMRON Digital Automatic Blood Pressure Monitor
:
HEM-907 (HEM-907-E7)
:
Digital display
:
Oscillometric method
:
Pressure; 0 to 299 mmHg
Pulse rate; 30 to 199 beats/min
:
Pressure; Within +/-3 mmHg
Pulse rate; Within +/-5 % of reading
:
Automatic inflation with pumping
:
Automatic deflation by electromagnetic control valve
:
Automatic rapid air release by electromagnetic control valve
8.0V 0.7A (AC adapter)
:
4.8V (Battery pack)
:
AC adapter (100 - 240 V , 50 - 60 Hz, 0.2 A)
or battery pack
:
Class II (AC Adapter)
Internal powered equipment (when operating with battery only)
:
Type BF (Applied part): Cuff
:
10 to 40 °C, 30 to 85 % RH (non-condensing)
:
-20 to 60 °C, 10-95 % RH (non-condensing), 700-1060 hPa
:
Approximately 910 g
:
139 (W) x 203 (H) x 131 (D) mm
:
Medium size adult arm cuff, AC adapter, air tube (1 m), Instruction
Manual (with guarantee card), Battery pack
:
Large size adult arm cuff, medium size adult arm cuff, small size adult arm cuff,
air tube (1.3 m), stand exclusive for this unit, wall-hanging kit, pole-mounting kit