Important safeguards/notices for use in the
medical environments
1. All the equipments connected to this unit shall be
certified according to Standard IEC60601-1,
IEC60950-1, IEC60065 or other IEC/ISO Standards
applicable to the equipments.
2. Furthermore all configurations shall comply with the
system standard IEC60601-1-1.
Everybody who connects additional equipment to
the signal input part or signal output part configures
a medical system, and is therefore, responsible that
the system complies with the requirements of the
system standard IEC60601-1-1.
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected
to other equipment.
4. For this particular equipment, all accessory
equipment connected as noted above, must be
connected to mains via an additional isolation
transformer conforming with the construction
requirements of IEC60601-1 and providing at least
Basic Insulation.
5. This equipment generates, uses, and can radiate radio
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference to other equipment. If this unit causes
interference (which can be determined by
unplugging the power cord from the unit), try these
measures: Relocate the unit with respect to the
Guidance and manufacturer's declaration-electromagnetic emissions
The LMD-2451MT is intended for use in the electromagnetic environment specified below.
The customer or the user of the LMD-2451MT should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Warning
If the LMD-2451MT should be used adjacent to or
stacked with other equipment, it should be observed to
verify normal operation in the configuration in which
it will be used.
Consult your dealer. (According to standard EN60601-
1-2 and CISPR11, Class B, Group 1)
6. Model LMD-2451MT is a monitor intended for use
Important EMC notices for use in the
medical environments
• The LMD-2451MT needs special precautions
• The portable and mobile RF communications
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold
by Sony Corporation, may result in increased emissions
or decreased immunity of the LMD-2451MT.
Compliance
The LMD-2451MT uses RF energy only for its internal
Group 1
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The LMD-2451MT is suitable for use in all establishments,
Class B
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class D
Complies
susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
in a medical environment to display pictures from
cameras or other systems, other than diagnostic X-
ray equipment.
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this instructions for use.
equipment such as cellular phones can affect the
LMD-2451MT.
Electromagnetic environment-guidance