Herunterladen Inhalt Inhalt Diese Seite drucken

Symbols On The Unit; Important Emc Notices For Use In The Medical Environments - Sony LMD-1530MD Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für LMD-1530MD:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Symbols on the unit

Symbol
Location
Front
1
Rear
Rear
Front
-

Important EMC notices for use in the medical environments

• The LMD-1530MD needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the LMD-1530MD.
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold
by Sony Corporation, may result in increased emissions
or decreased immunity of the LMD-1530MD.
The LMD-1530MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the LMD-1530MD should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Warning
If the LMD-1530MD should be used adjacent to or
stacked with other equipment, it should be observed to
verify normal operation in the configuration in which
it will be used.
30
This symbol indicates
Standby switch.
The equipotential
terminal which brings the
various parts of a system
to the same potential.
Functional earth terminal
Key inhibit
The setting are locked so
that they cannot be
changed.
Guidance and manufacturer's declaration-electromagnetic emissions
Compliance
Group 1
Class B
Class D
Complies
This symbol is located on the rear of the unit.
See page 42 of these instructions for details
about how to remove, or attach the connector
cover.
Refer to the operating instructions
Follow the directions in the operating
instructions for parts of the unit on which this
mark appears.
This symbol indicates the manufacturer, and
appears next to the manufacturer's name and
address.
Electromagnetic environment-guidance
The LMD-1530MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The LMD-1530MD is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis