BATTERIES RC
RED light
C A U S E
S O L U T I O N
Problem: the lift does not travel in either direction and the LED on the remote
control does not light up when the remote control is operated.
Cause: the batteries in the remote control are flat.
Solution: replace the batteries in the remote control.
POWER
45
15
15 sec.
30
C A U S E
S O L U T I O N
Problem: the lift batteries are not recharging, the lift beeps for 15 seconds and
the display shows a "2".
Cause: the lift is not in the correct position at a charge point or the plug is not in
the socket.
Solution: move the lift to a charge point or put the plug in the socket.