Technical Information
Window Meanings

Soft Start

"Soft Start" mode is provided to start the pump after a protracted
stop of the Turbo-V 2300 TwisTorr Pump.
This allows a better grease distribution in the bearings.
The "Soft Start" mode is disabled by default. The "Soft Start" mode
can be activated by a suitable software, by Remote I/O or by Front
Panel command (see the paragraphs "INTERCONNECTIONS" and "RS
232/485 COMMUNICATION DESCRIPTION"). "Soft Start" frequency
steps are the followings: 200 Hz, 225 Hz, 250 Hz, 275 Hz, 300 Hz and
555 Hz. The pump remains at each step for a period of 300 s. A fail
condition occurs if the pump is not able to properly spin up the
driving frequency (Too High Load Error).