B. Braun Aesculap Spine activ C FW857R Gebrauchsanweisung Seite 9

Einsetzinstrument
Vorschau ausblenden Andere Handbücher für Aesculap Spine activ C FW857R:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
Manual cleaning with immersion disinfecting and cleaning with brush
Stage
Step
I
Disinfecting
cleaning
II
Intermediate rinse
III
Final rinse
IV
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized)
RT:
Room Temperature
Stage I
Fully immerse the product in the disinfecting
solution. Make certain that all accessible surfaces
are moistened.
Clean the product with a suitable cleaning brush
until all visible residues have been removed from
the surface.
Brush through all surfaces that are not accessible
to visual inspection, e.g. in products with hidden
crevices, lumens or complex geometries, for at
least 5 min or until no more residues can be
removed. Mobilize non-rigid components, e.g. set
screws, links, etc. during cleaning.
After that, thoroughly rinse (at least 5 times) with
the cleaning solution these difficult to access areas
of the product, using a single-use syringe (20 ml).
To avoid the risk of corrosion, do not use a metal
brush or other abrasives, which would damage the
surfaces, for cleaning the product.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
15
2
(cold)
RT
1
-
(cold)
RT
0.5
-
(cold)
RT
-
-
Stage II
Stage III
Stage IV
Water
quality
D–W
BBraun Stabimed; aldehyde phenol
and QAV free; pH = 9
D–W
-
FD–W
-
-
-
Rinse the product completely (all accessible
surfaces) under running water.
Carry out a full rinse of the product (all accessible
surfaces) under running water.
Allow water to drip off for a sufficient length of
time.
Dry the product completely with compressed air.