LiNA OperåScope™ Tubing & Drape Kit
4. Carefully inspect the packaging for any
damages prior to use. Do NOT attempt to use
the device if the sterile barrier is damaged. Do
NOT use if past expiration date.
5. Remove the inner pouch from the sterilized
blister (Figure 1).
6. Remove the drape from the inner pouch
(Figure 2a).
7. Inspect the drape for labels for left & right
hand. Unfold the drape by inserting the left &
right hand at the labels, and pull them away
from each other (Figure 2b-2d).
8. Place the drape between the gynecological
bench and the patients' buttock. Let the pouch
hang down towards the floor (Figure 2e).
9. Remove the collection bag from the inner
pouch. Close the drain valve at the bottom
and remove the protection cap from the inlet
(Figure 3a).
10. Remove the OUTflow tubing from the inner
pouch. Connect the luer to the OUT connector
on LiNA OperåScope™ and connect the other
end to the collection bag or put it into the
drape pouch (Figure 3b – 3e).
11. Remove the INflow tubing from the inner
pouch. Connect the luer to the IN connector on
LiNA OperåScope™ (Figure 3f).
12. Remove the protection cap from the spike and
insert the spike into the saline bag (Figure 3g).
The slide clamp may be used to temporarily
stop the flow from the saline bag, if needed.
13. Apply light pressure to the saline bag using the
pressure cuff.
14. Open the inlet luer on LiNA OperåScope™
and flush fluid through the tubing and LiNA
OperåScope™ until all air bubbles have cleared
from the tubing. Open the outlet luer.
15. Insert the LiNA OperåScope™ into the patient
and perform the procedure.
16. Adjust the pressure on the saline bag and
outlet cock position to distend the uterus.
17. Throughout the procedure, monitor the fluid
deficit to avoid fluid overflow (Figure 4).
18. When the procedure is complete, remove the
LiNA OperåScope™ from the patient and close
the inlet luer, remove the pressure from the
saline bag. Disconnect the tubings from the
LiNA OperåScope™, saline bag and collection
pouch.
19. Dispose of the fluids from the saline bag, the
collection bag, and the drape pouch. If needed,
cut the drain plug on the pouch (Figure 5).
LiNA Medical ApS - Formervangen 5 - DK-2600 Glostrup - DENMARK - Tel: +45 4329 6666 - Fax: +45 4329 6699 - e-mail: info@lina-medical.com - www.lina-medical.com
Ref: OP-TDK-6-I
DISPOSAL
After use, the product may be a potential
biohazard. Handle and dispose the fluids, drape,
tubings and collection bag in accordance with
local, state and federal laws and regulations
(Figure 5).
Country of origin: Poland.
CE marking 2020.
LiNA is a registered trademark of LiNA Medical in EU, US and CN
en
2021-11 FV0279B