1000 Series
The calibration menu is only shown when the scale is not in LFT/Approved mode
before verification and sealing. Refer to section 5. Legal For Trade for more
information.
SPAN
SPAN [
]
Span calibration uses two calibration points, one at zero load and the other at
specified full load (span). For detailed calibration mass information, please refer to
the table in section 2.4.
With no load on the balance, enter the Span Calibration to begin the process.
-
Once the scale displays
and press Yes again to perform a Span calibration.
-
After the scale shows
the screen. Place weight(s) of shown value and press Yes.
-
Then the scale shows
successfully, the scale displays
automatically.
If incorrect calibration masses are used, the scale displays
user menu automatically.
To abort calibration at any point, press No.
LIN
Lin [
]
When performing a linearity calibration, the scale will prompt you for a zero weight -
- the platform without anything on it and weight display at zero -- a Midpoint weight
equal to 50% of the Span weight, and a Span weight equal to the full capacity of the
scale.
Maximum
Capacity (kg)
2 kg
3 kg
5 kg
6 kg
10 kg
15 kg
20 kg
30 kg
With no load on the balance, enter the Linearity Calibration to begin the process.
-
After entering the calibration menu, press No to move to linearity
calibration menu
C.A.L
--C--
--C--
Maximum
Capacity (lb)
5 lb
6 lb
10 lb
15 lb
20 lb
30 lb
50 lb
60 lb
LIN
and press Yes to start a linearity calibration.
, press Yes to enter the calibration menu
, the calibration mass value will be shown in
. Once the span calibration is completed
Done
and exits the user menu
Linearity Calibration
Points (kg)
0, 1 kg, 2 kg
0, 1.5kg, 3 kg
0, 3kg, 5 kg
0, 3kg, 6 kg
0, 5kg, 10 kg
0, 10kg, 15 kg
0, 10kg, 20 kg
0, 15kg, 30 kg
Cal E
and exits the
Linearity
Calibration
Points (lb)
0, 3lb, 5 lb
0, 3lb, 6 lb
0, 5lb, 10 lb
0, 10lb, 15 lb
0, 10lb, 20 lb
0, 15lb, 30 lb
0, 30lb, 50 lb
0, 30lb, 60 lb
EN-9