8. Calibration functions

The TK-1000 enables the output of different signal sources for simulation of different
temperature sensors. Below you will find descriptions of various functions.
a) Voltage output "mV"
The "Voltage Output" mode enables to select two DC voltage ranges.
0 - 100 mV with a resolution of
0.001 mV and
0 - 1000 mV with a resolution
of 0.1 mV.
Switch on the device and wait
until the operating display ap-
pears on the screen.
To switch to the 100 mV range,
press the "V/Ω" button (3) until
the desired range is displayed.
The range advances with each
keystroke (100 mV -> 1000 mV -> 500 Ω -> 5000 Ω -> 100 mV ...). The voltage value can be
set directly.
The four cursor buttons (5, 6, 12, 13) can be used to set the decimal place and the value. A point
or decimal place advances by briefly pressing the respective button.
The values can be set quickly by pressing and holding the cursor buttons for >1.5 s. The respec-
tive % values are called up (0%, 100% or in 25% increments).
0 - 100 mV
0 - 5000
0 - 1000 mV
0 - 500