7.6 Parts counting
Switch-on the balance using the
Press
key shortly.
The reference unit number 5 appears
By pressing the
key several times other reference quantities 10, 25 and 50 can
be called up.
Place as many parts to count on the weighing plate as the set reference quantity
requires.
Confirm with
key.
The balance is now in parts counting mode counting all units on the weighing plate.
By pressing the
key the balance returns to the weighing mode and displays the
weight of the counted units.
Important: The larger the reference quantity, the more accurate the parts
counting.
Smallest counted weight see table „Technical data", if this weight is less, in the
display appears „Er 1". Use
The tare vessels can also be used for piece counting. Before starting percent
weighing use the
PCB-BA-e-0712
key and wait for the „0" display.
key to return to the weighing mode.
key to tare out the container.
17