Herunterladen Inhalt Inhalt Diese Seite drucken

Selection Of The Adjustment Weight - KERN PCB Serie Betriebsanleitung

Präzisionswaage
Vorschau ausblenden Andere Handbücher für PCB Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 182

8.4 Selection of the adjustment weight

In the model KERN PCB, the adjustment weight can be selected from four pre-set
nominal values (approx.1/4; 1/2; 3/4; max) (refer also to table 1 below, factory setting
with grey background). In order to achieve high-quality weighing results in the sense
of the measuring technology, it is recommended to select the nominal value as high
as possible.
Table 1:
PCB 40-3
PCB 60-3
10g
20g
30g
40g
PCB 400-1
PCB 400-2
100g
200g
300g
400g
PCB1000-2
PCB 1600-2
200g
500g
1000g
700g
1200g
1000g
1500g
PCB 6000-0
1000g
2000g
5000g
6000g
PCB-BA-e-0712
PCB 100-3
10g
20g
20g
50g
50g
70g
60g
100g
PCB 600-2
100g
100g
200g
200g
300g
400g
400g
500g
600g
PCB2000-1
500g
500g
1000g
1500g
2000g
PCB 6000-1
1000g
2000g
5000g
6000g
PCB 160-3
50g
100g
120g
150g
PCB 800-2
200g
500g
700g
800g
PCB 4000-0 PCB 4000-1
1000g
2000g
3000g
4000g
PCB 8000-1
PCB 10000-1
2000g
5000g
7000g
8000g
PCB 200-2
50g
100g
150g
200g
PCB1000-1
200g
500g
700g
1000g
1000g
2000g
3000g
4000g
2000 g
5000g
7000g
10000g