Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 48 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 48 should
assure that it is used in such an environment.
IMMUNITY test IEC 60601
test level
3 V/m 80
Radiated RF
MHz to
IEC 61000-4-3
2.5 GHz
All manuals and user guides at all-guides.com
C o m p l i -
Electromagnetic environment – guidance
ance level
Portable and mobile RF communications equipment should be used no closer to
any part of the BM 48, including cables, than the recommended separation dis-
tance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d
P
1
2 .
d
P
2
3 .
Where P is the maximum output power rating of the transmitter in watts (W) ac-
3 V/m
cording to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,
Interference may occur in the vicinity of equipment
marked with the following symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
should be less than the compliance level in each frequency range
114
.b