NOTE:
Do not use a brushes or any other tools
to clean the aerosol head as they may
cause damage to the product
components.
• Rinse all nebuliser handset parts and
the front and back of the aerosol head
thoroughly under running tap water.
• Shake all parts to remove excess water
more quickly.
• Place the individual parts on a dry, clean
surface and allow them to air dry
completely.
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
each cleaning.

Disinfection

Disinfect the nebuliser handset parts and
the aerosol head at least once a day, e.g.,
immediately after cleaning at the end of
the day (a nebuliser handset cannot be
®
eFlow
rapid - 2021-04
disinfected effectively unless it has been
cleaned first).
NOTE:
Do not use a microwave to disinfect the
aerosol head. The possibility of product
components being damaged cannot be
ruled out.
With a commercially available
disinfector for baby bottles
• To ensure effective disinfection, use
a thermal disinfector with an operating
time of at least 6 minutes. With regard to
performing the disinfection, the duration
of the disinfection process and the
dosing of water required, follow the
instructions for use of the disinfector you
are using.
CAUTION:
Inadequate disinfection promotes the
growth of bacteria and thus increases
the risk of infection. If disinfection is not
carried out for the time specified for the
device concerned, it will not be effective.
Therefore, do not switch the device off
before the disinfection cycle is com-
plete. Also, make sure that the device is
clean and check it regularly for proper
operation.
en
51