Battery life
For approx. 300 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA bat-
teries, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may
interfere with this unit. More details can be requested
from the stated Customer Service address or found at the
end of the instructions for use.
• This device is in line with the EU Medical Devices Direc-
tive 93/42/EC, the "Medizinproduktegesetz" (German
Medical Devices Act) and the standards EN 1060-1
(non-invasive sphygmomanometers, Part 1: General
requirements), EN 1060-3 (non-invasive sphygmoma-
nometers, Part 3: Supplementary requirements for elec-
tro-mechanical blood pressure measuring systems)
and IEC 80601-2-30 (Medical electrical equipment –
All manuals and user guides at all-guides.com
• The accuracy of this blood pressure monitor has been
26
Part 2 – 30: Particular requirements for the safety and
essential performance of automated non-invasive blood
pressure monitors).
carefully checked and developed with regard to a long
useful life. If using the device for commercial medical pur-
poses, it must be regularly tested for accuracy by appro-
priate means. Precise instructions for checking accuracy
may be requested from the service address.