Herunterladen Inhalt Inhalt Diese Seite drucken

Endress+Hauser Micropilot FMR20 Modbus RS485 Kurzanleitung Seite 55

Freistrahlendes radar
Vorschau ausblenden Andere Handbücher für Micropilot FMR20 Modbus RS485:
Inhaltsverzeichnis

Werbung

Micropilot FMR20 Modbus RS485
Adresse
Name
5280
LCRS_ADJU
STEVENTE-
CHOLOST
5281
LCRS_CHAN
GINGVELO-
CITY
5282
LE_CSTLIN-
TYPE
5283
LE_CUSTO-
MUNIT
5412
LCRS_EVA-
LUATIONS-
ENSITIVITY
5413
LE_LEV-
CORROFFS
5415
LE_CT_ACT
TABLE
5416
LE_OUT-
PUTMODE
5417
LE_CT_EDIT
MODE
5418
LCRS_EVA-
LUATION-
DISTANCE
5420
LCRS_FREE-
FIELDMODE
5421
LCRS_HIGH
BLOCKDIS-
TANCE
Endress+Hauser
Anzahl
Zugriff
Register
• Service
• Bediener
1
• read/
write
• read only
1
• read/
write
• read only
1
• read/
write
• read only
1
• read/
write
• read only
1
• read/
write
• read only
2
• read/
write
• read only
1
• read/
write
• read only
1
• read/
write
• read only
1
• read/
write
• read only
2
• read/
write
• read only
1
• read/
write
• read only
2
• read/
write
• read only
Bereich
Datentyp
warning: 0
enum8
alarm: 1
>1m/min: 0
enum8
<1m/min: 1
<10cm/min: 2
no Filter/Test:
3
not active:
enum16
32989
active: 33171
see Parameter
enum16
Einheit nach
Linearisie-
rung
low: 0
enum8
medium: 1
high: 2
–25 ... 25
float32
disable: 32852
enum16
enable: 32887
level: 32949
enum16
empty: 33197
edit manual:
enum16
32890
delete table:
32889
0 ... 21,8
float32
off: 33004
enum16
on: 33006
0 ... 20
float32
Modbus Parameter Übersicht
SI-Einheit
Beschrei-
bung
-
Defines the
lost echo
status
-
Defines
maximum
change velo-
city of an
echo change
-
Is lineariza-
tion table is
used
-
Customer
level unit
-
Sensitivity of
the echo
detection
-
Level correc-
tion
-
States, if
linearization
table is
active
-
States the
output mode
-
Edit mode of
the linear-
ization table
m
Max dis-
tance which
will be eva-
luated
-
Switch to
free field
parameters
m
Blocking dis-
tance (dis-
tance from
sensor which
is not eva-
luated)
55

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis