Herunterladen Inhalt Inhalt Diese Seite drucken

Storz E-Box TP012 Gebrauchsanweisung Seite 18

Inhaltsverzeichnis

Werbung

16
Anhang
EMV
The E-Box TP012 is intended for use in the electromagnetic environment specified below. The user of E-Box TP012 should make sure that it is used in
such an environment.
Immunity tests
EN/IEC 60601 test level
Electrostatic discharge (ESD)
± 8 kV Contact discharge
acc. to IEC 61000-4-2
± 15 kV Air discharge
Rapid transient electrical
± 2 kV for power lines
Interferences/bursts acc. to
± 1 kV for input and output lines
IEC 61000-4-4
100 kHz repetition
Surges
± 1 kV line(s) to line(s)
acc. to IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
Voltage Dip:
interruptions and fluctuations
Dip to 0 % for 1 cycles
of the supply voltage
@ 0° phase angle
acc. to IEC 61000-4-11
Dip to 70 % for 25/30 cycles
@ 0° phase angle
Dropout to 0 % for 0.5 cycles
@ 0°, 45°, 90°, 135°, 180°,
225°, 270° & 315° phase angles
Voltage interruption:
100 % for 250/300 cycles
Magnetic field at the supply
30 A/m at 50 Hz/60 Hz
frequency
(50/60 Hz)
acc. to IEC 61000-4-8
Radiated, radio-frequency,
3 V/m 80 MHz to 2.7 GHz
electromagnetic field immunity
*Refer table 2 for wireless Proximity
test acc. to IEC 61000-4-3
RF field test levels
Immunity to conducted
3 V
on 150 kHz to 80 MHz
rms
disturbances, induced
1 kHz 80% AM modulation
by radio-frequency fields
6 V
 in ISM bands
rms
acc. to IEC 61000-4-6
Appendix
EMC
Table 1 – Compliance level for Immunity tests
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
± 8 kV Contact discharge
± 15 kV Air discharge
± 2 kV/1 kV
± 1 kV for input and output lines
100 kHz repetition
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage Dip:
Dip to 0 % for 1 cycles
@ 0° phase angle
Dip to 70 % for 25/30 cycles
@ 0° phase angle
Dropout to 0 % for 0.5 cycles
@ 0°, 45°, 90°, 135°, 180°, 225°,
270° & 315° phase angles
Voltage interruption:
100 % for 250/300 cycles
30 A/m at 50 Hz/60 Hz
3 V/m 80 MHz to 2.7 GHz
3 V
on 150 kHz to 80 MHz
rms
1 kHz 80 % AM modulation
6 V
 in ISM bands
rms
Anexo
CEM
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the E-Box TP012 require continued
operation during power mains interruptions, it is
recommended that the E-Box TP012 be powered
from an uninterruptible power supply or a battery.
If image distortion occurs, it may be necessary to
position the E-Box TP012 further from sources
of power frequency magnetic fields or to install
magnetic shielding. The power frequency magnetic
field should be measured in the intended installation
location to assure that it is sufficiently low.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis