Installing the ferrite bead for the battery
1. Place the ferrite bead for the battery around the battery's
Fit capacitor for charging interface
1. Loosen and remove cover panel.
2. Use a size 5 wrench to stop the contact pins turning and undo
3. Pass the cable lugs for the capacitor to the outside. Use the
4. Put the cable lugs for the capacitor on the contact pins for the
5. Wet the threads of the contact pins with a drop of Loctite 222
6. Place washers and nuts on the threads.
6
EN
Accuvac Rescue
connecting cables and squeeze it together until it clicks.
the nuts using a size 7 ring wrench.
same slots through which the cables to the charging interface
pass.
charging interface. In the process, please note: place cable lugs
of identical color on top of one another (red on red, black on
black).
(WM 14958 or WM 14915).
Size 7
Size 5
Cable lugs