Specifications

2
Specifications
2.1

Measured values

Differential Pressure
Internal Temperature
Measurement
External Temperature
Measurement
(e.g. with temperature measuring
clamp )
22
Description
Default unit
Range
Permissible overload
Precision
Resolution
Description
Default unit
Range
Precision
Resolution
Description
Range
Precision
Resolution
Data
Pa, mbar or inWC,
depending on device
version
± 3.000 hPa
± 3.200 hPa
< 3 % of reading,
when < ± 2 hPa
better than ± 0,06 hPa
0,01 hPa
when –210 hPa to
+210 hPa,
otherwise 0,1 hPa
Data
°C or °F, depending
on device version
-20 °C to 60°C
< ± 1 °C
0,1°C
Data
-20,0 °C to +800,0 °C
< ± 1 °C
when -20°C to 67°C,
otherwise 1,5 % of reading
according to EN 50379-1