maximum field strength of the source of interfer-
ence. In this way it is possible to, for example, trace
cables in plaster, provided these are live, independ-
ent of whether a load is connected to the cable or
not.
Caution!
Particularly during the measurement of E-fields, the
human body may have a significant effect on the
field. That is, contact with the device or the sensor
can distort the measurement.
To determine precise measured values, if neces-
sary the meter and the sensor should be placed
close together and not touched.
If you only want to use the meter to track down
sources of interference, it is not necessary to fol-
low this instruction, as in this case it is initially the
location of the peak value that is of interest.

Changing the measuring range

By briefly pressing the "E/H" button you can change
between the measuring ranges 0 to 200 V/m and 0
to 2 kV/m. It is only possible to change to the lower
measuring range if the current measured value does
not exceed the maximum value for this measuring
range (200 V/m). The low measuring range is indi