Maintaining the product
Replacing a a v v alve:
Unlock valve knob and open completely.
1
First pull out the two lateral socket covers, then remove
2
the protective cover.
Press lightly on the valve knob so that it is prevented
3
from jumping out.
Release the locking bracket by lifting and remove it.
4
Attention: The locking bracket consists of two
components (bracket
Remove valve knob with spring.
5
Unscrew valve screw (wrench size 14mm) and remove
6
valve (consisting of valve piston
If necessary exchange valve piston and valve screw and
7
replace the parts. Torque: 4.0Nm / 2.95ft*lb.
Place the valve knob over the valve and turn until the
8
knob can be pushed into the housing.
Press the valve knob lightly into the housing and clip on
9
the locking bracket (bracket and slide).
Place the instrument in the protective cover and press
10
the edges into the clip fastening, then replace the two
lateral socket covers (observe left / right).