Control Solution Testing
Running a control test lets you know that your meter and test
strips are working properly to give reliable results. You should
run a control test when:
• You use the smartLAB
first time.
• You open a new vial of test strips.
• You think the meter or test strips might be working incorrectly.
• You drop the meter.
• You have repeated a test and the test results are still lower or
higher than expected.
• You are practicing the test procedure.
Note: Professional users are instructed to follow federal, state,
and local guidelines concerning QC practices.
About the smartLAB
•
smartLAB
consumption or injection.
Use only with smartLAB
•
•
Write the date you first opened the bottle on the bottle
label.
The smartLAB
•
months from the date the vial is opened or until the „Expira-
tionDate"onthebottle,whichevercomesfirst.
Do not use smartLAB
•
sprint nG Blood Glucose System for the
®
nG control solution
®
nG control solution is not intended for human
®
nG control solution is durable for three
®
nG control solution that is past the
®
nG test strips.
®
62