Control Solution Testing
Running a control test lets you know that your meter and test strips
are working properly to give reliable results. You should run a con-
trol test when:
• You use the smartLAB
first time.
• You open a new vial of test strips.
• You think the meter or test strips might be working incorrectly.
• You drop the meter.
• You have repeated a test and the test results are still lower or
higher than expected.
• You are practicing the test procedure.
Note: Professional users are instructed to follow federal, state, and
local guidelines concerning QC practices.
About the smartLAB
• smartLAB
consumption or injection.
• Use only with smartLAB
• Write the date you first opened the bottle on the bottle label.
• The smartLAB
from the date the vial is opened or until the „Expiration Date"
on the bottle, whichever comes first.
• Do not use smartLAB
„Expiration Date"
®
sprint nG Blood Glucose System for the
nG control solution
®
nG control solution is not intended for human
®
®
nG control solution is durable for three months
nG control solution that is past the
®
®
nG test strips.
62