3. The icons
show themselves.
4.Placethemeteronaflatsurface,like
a table.
5. Remove the control solution bottle
cap and wipe the tip of the bottle with
a tissue.
6. Squeeze the bottle until a tiny drop
forms at the tip of the bottle.
7. Touch the drop to the Blood collection
area at the tip of the test strip. Do not
put control solution on top of the test
strip.
8. The meter starts to count down from 5
9. Do not remove the test strip yet. Check if the
reading falls within the range printed on the test
strip bottle.
10. Push the Strip Ejector to remove the test strip.
11. Throw the test strip away after you have com-
pared the reading to the range printed on the test
strip bottle.
Note: Be aware that there are Normal and High ranges listed
on your test strip bottle. The Normal and High refer to the two
different control solutions that are available. Please be sure that
you are looking at the correct range.
(insert strip) and
(apply blood)