EMC Tables 204 and 206
Immunity test
IEC 60601 test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to < 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the unit is used exceeds the applicable RF compliance level above, the unit
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the display.
Rated maximum
output power of
transmitter
[W]
Guidance and manufacturer's declaration – electromagnetic immunity –
for equipment and systems that are not life-supporting
Compliance level
3 volts
3 V/m
Separation distance d [m] according to frequency of transmitter
150 kHz to < 80 MHz
3
5 .
=
d
P
⎢ ⎣
⎥ ⎦
3
0.0
0.1
1.0
10.0
100.0
Table 204
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
[ ]
3
5 .
=
d
P
3
[ ]
3
5 .
=
d
P
3
[ ]
7
=
d
P
800 MHz to 2.5 GHz
3
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range
Interference may occur in the vicinity of equipment marked with the
following symbol:
Table 206
80 MHz to < 800 MHz
3
5 .
=
d
P
⎢ ⎣
⎥ ⎦
3
0.12
0.38
1.2
3.8
12
150 kHz to < 80 MHz
80 MHz to < 800 MHz
a
, should be less than the compliance
b
.
800 MHz to 2.5 GHz
7
=
d
P
3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
18