Spencer supplies accessories for operating in this lifting situation.
Always verify correct fastening between the harness and the stretcher fixing point. For instructions on how to use the harnesses, check the relevant product-specific user
manual and also comply with the provisions of this manual.
Before lifting or lowering, after carrying out all the safety checks necessary for handling, adjust the harness and/or distribute the load inside the basket stretcher in order
to balance its horizontal position when it is then handled.
Whenever a patient is lifted onto a basket stretcher, the patient should be secured to the stretcher using an appropriate number of belts.
Using the STX 519 system

12. CLEANING AND MAINTENANCE

Spencer Italia S.r.l. declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product and spare parts and/or in any
case of any repair carried out by a person other than the Manufacturer, who uses internal and external technicians authorised to do so; moreover, doing so will invalidate
IT
the warranty.
• The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking, maintenance and cleaning operations.
• Establish a maintenance schedule, periodic inspections and extend the average life span, if foreseen by the Manufacturer in the User Manual, identifying a reference person
who meets the basic requirements set forth in the User Manual.
• The frequency of inspections is determined by factors such as legal requirements, type of use, frequency of use, and environmental conditions during use and storage.
• Repairs of products manufactured by Spencer Italia S.r.l. must be carried out by the Manufacturer, who shall make use of specialised internal or external technicians
who, using original spare parts, shall provide quality repair service   in strict compliance with the technical specifications indicated by the Manufacturer. Spencer Italia S.r.l.
declines any responsibility for any direct or indirect damage which is a consequence of improper use of spare parts and/or any repair work carried out by unauthorised
parties.
EN
• All maintenance and overhaul activities must be recorded and documented with the relevant technical operation reports. This documentation must be kept for at least 10
years after the end of the product's life and must be made available to the competent authorities and/or the Manufacturer when requested.
• Cleaning, provided for reusable products, must be carried out in accordance with the Manufacturer's instructions in the User Manual in order to avoid the risk of cross-in-
fection due to the presence of body fluids and/or residues.
• If required, the product and all its components must be washed and left to dry completely before storage.

12.1 CLEANING

Failure to carry out the correct cleaning operations could increase the risk of cross-infection due to presence of body fluids and/or residues.
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking and cleaning operations.
DE
• Clean the exposed parts with water and delicate soap. Never use solvents or stain removers.
• Rinse thoroughly with lukewarm water, making sure you have removed all traces of soap, which may deteriorate or compromise conditions and durability. Avoid using high
pressure water, as it penetrates the joints and removes lubricant, creating the risk of corrosion on components. Let dry completely before storing. Drying after washing or
after use in a wet environment must be natural and not forced. Do not use flames or other direct heat sources.
• After complete drying, proceed with lubrication as described below.
• If disinfecting, use products that do not have a solvent or corrosive action on materials constituting the device, in addition to being classified as medical-surgical devices.
• Carefully follow the instructions of the manufacturer of the product used with regards to the application method and contact time.
• Be sure to take all precautions to ensure that there is no risk of cross-infection or contamination of patients and operators.
FR

12.2 ROUTINE MAINTENANCE

You must establish a maintenance program and periodic checks, identifying a designated reference person. The person entrusted with device maintenance must ensure
the basic requirements envisaged within this user manual.
All routine and special maintenance activities and all general overhauls must be recorded and documented with the relevant technical intervention reports. This
documentation must be kept for at least 10 years after the end of the device's life and must be made available to the competent authorities and/or the Manufacturer
when requested.
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking, maintenance and cleaning operations.
ES
THE MAINTENANCE SCHEDULE MUST COMPLY WITH THE FOLLOWING TABLE:
Minimum maintenance intervals
Disinfection
Cleaning
Lubrication
Inspection
PT
THE INSPECTION TO BE CARRIED OUT AFTER EACH USE INVOLVES:
• Check that all components are present
• Check that the belts are in good condition without cuts, burns, abrasions, open seams or fraying
• Check that the seams are in good condition, with no fraying or breakage
• Check that rings and spring catches comply with their original shape, are free of cracks, deep scratches and cutting Extremely careful analysis to determine if what looks
like a scratch is actually a crack
• Check that the moving parts slide properly
• Check that the spring catch pins are secured in place
• Check that the spring catch locking sleeves (if present) rotate correctly and allow for secure locking
• Check that the spring catch springs are able to keep the spring catch tightened securely
• Check that the metal components are not oxidised
• Check of the readability of labels for warnings, data, life span and range
• General check of the state of wear of each component
• Disinfection – Par. 12.1
14
Using the STX 540 system
At each use
Using the STX 519 system for dragging
If necessary