Herunterladen Inhalt Inhalt Diese Seite drucken

M&C MP Serie Betriebsanleitung Seite 51

Vorschau ausblenden Andere Handbücher für MP Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Diaphragm Pumps PJ19423-726 EX
Translation of Original Operating- and Installation Instructions, english, KNF PJ19423-726 EX 09/16 21
2. Position the pump head on the housing A according to the felt-
tip pen marking.
3. Fit the 4 allen screws H and tighten them uniformly diagonally
until the disc springs R are flattened.
To prevent parts from warping, start by tightening each of the
screws half a turn alternately.
4. At the valve bodies F and G tighten each three allen screws D
carefully until resistance is noticeable.
5. Check the smooth movement of the pump by turning the
counterweight P.
6. Mount cover N with the four cheese head screws M.
7. Check and ensure the gas-tightness of the pump head and the
tubing:
Hazard of explosion caused by leaks
 Before putting the pump into operation again,
check and ensure the gas-tightness of the pump
head and the tubing. Leakages may lead to ex-
WARNING
plosion hazard!
Servicing

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mp47-z ex02p1250

Inhaltsverzeichnis