Anhang 1

State
Low voltage alarm
SpO
alarm
2
Pulse rate alarm
Probe error alarm
Anhang 2
Guidance and manufacturer's declaration – electromagnetic emission
The RS01 is intended for use in the electromagnetic environment specified below. The customer of the user of the RS01 should assure that it is used in
such and environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Guidance and manufacturer's declaration – electromagnetic immunity
The RS01 is intended for use in the electromagnetic environment specified below. The customer or the user of RS01 should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
NOTE
The RS01 is intended for use in the electromagnetic environment specified below. The customer or the user of RS01 should assure that it is used in such
an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the RS01 is used exceeds
the applicable RF compliance level above, the RS01 should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the RS01.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The RS01 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the RS01
Alarm condition delay
1s
330ms
330ms
16ms
Guidance and manufacturer's declaration – electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
IEC 60601 test
Compliance level
level
6 kV contact
6 kV contact
8 kV air
8 kV air
3A/m
3A/m
Guidance and manufacturer's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
Compliance level
3 V
3 V
rms
rms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM –
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the RS01
Alarm signal generation delay
20ms
20ms
20ms
20ms
Compliance
Electromagnetic environment – guidance
The RS01 uses RF energy only for its internal
function. Therefore, its RF emissions are very
Group 1
low and are not likely to cause any interference
in nearby electronic equipment.
The RS01 is suitable for use in all establishments
other than domestic and those directly
Class B
connected to a low voltage power supply
network which supplies buildings used for
domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floor
are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the RS01, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
3
5 .
d
P
V
1
3
5 .
d
P
E
1
7
d
P
E
1
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
24
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
should be less than the
b