Step 3
Do not press the power button when the "hour glass" is
on the screen!
Step 4
Firmware updated successfully, OK to turn power off and remove SD card.
11.
Cleaning Instruction
1.
Alcohol wipes or mild detergent and water to remove dirt and grease from the product.
2.
Recommend soft optical cloth for cleaning the LCD window.
3.
Recommend cotton cloth for cleaning imager head.
4.
Allow appropriate drying time before operating handheld device.
12.
Storage
1.
It is good practice to clean the unit prior to long term storage.
2.
It is best to store the unit in a cool dry location.
13.
Troubleshooting
Symptom
Display unit does not turn on
Display turns on, but does not
show image
Display shows an
unresponsive image
Battery does not hold a
charge
36
The firmware is updating (45 seconds).
Possible Reason
Battery is discharged
Faulty electronics
Loose cable connection
Imager head covered by
debris
Image processor encounteres
an error
Battery may have discharged
slightly during storage
Run time dependent on LED
usage
Battery may be damaged
Life cycles of battery may
have expired
ENGLISH
Solution
Recharge battery
Contact ROTHENBERGER
Service Center
Check cable connections
Clean imager head
Turn OFF unit/turn ON to
reset processor
Recharge battery
Recharge battery more often
when extensive use of LEDs
Contact service center and
replace battery
Contact service center and
replace battery