6. Assembly
6.7 Analyzer and Polarizer*
Analyzer
If the analyzer was inserted into the tube mount
before the tube assembly: (→ p. 19), no additio-
nal assembly step is required.
If an intermediate tube pole* or analyzer mount
TL* is used:
• Remove the plug cap on the left side.
• Insert the analyzer into the receptacle until it
latches in place.
Polarizer
• Raise the condenser to its upper stop position.
• Remove the DLF filter magazine from the base
if present.
• Press the polarizer holder in place (Fig. 18).
• Push the polarizer with the labeled side
upward into the lower opening.
Fig. 18
Filter holder
with 2 positions
24
Alternative:
• Attach the polarizer holder to the underside of
the condenser holder with the left clamp
screw (19.1). Remove the flip-out blue filter if
required.
• Push the polarizer with the labeled side
upward into the lower opening.
6.8 Lambda Plate Compensator*
• Raise the condenser to its upper stop position.
• Remove the DLF filter magazine from the base
if present.
• Attach the lambda plate compensator to the
base.
Fig. 19
Assembly of polarizer holder
1
Clamping screw
1