Herunterladen Inhalt Inhalt Diese Seite drucken

Gima SUPERVEGA EVO Handbuch Seite 21

Absauger
Vorschau ausblenden Andere Handbücher für SUPERVEGA EVO:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 42
21
• The negative pressure is set with keys - and +, which are found below the LCD screen: Press one of the
two keys with the motor on. The level of the negative pressure will flash at the centre of the display.
Release the key and, after 3 seconds, the value will be operating (fixed digit on the display) and saved. The
desired value remains fixed until the operator varies the negative pressure, whilst the bar-graph moves
along the semi-circumference stopping in correspondence (graph) of the value set.
The maximum negative pressure value that can be set is: -75kPa (-0.75 bar)
The minimum negative pressure value that can be set is: -15kPa (-0.15 bar)
• To suspend and / or terminate the treatment press the ON / OFF button and and pull the plug out from the
power socket.
• Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an
rapid reaching of the functionally vacuum) then rescrew the lid on the jar correctly.
• To extract the accessories and start with cleaning.
• At the end of each use place the device back in the box, protected against dust.
WARNING: The power cord plug is the element of separation from the electrical mains, even if the
device is equipped with an on/off button. Once the device is in use, the power plug must remain accessible to
allow another method of disconnection from the electrical mains.
The device features an energy saving function which allows the device to increase its autonomy and de-
crease the number of revs of the motor. In order to do so, it controls the negative pressure which is generated
only when the operator suctions body fluids.
If the operator does not carry out the suction process the device lowers the number of revs of the motor, thus
protecting the duration of the internal battery.
"Proximity" sensor
LCD display
Adjustment button
depression "-"
Operation with internal battery
• Connect the device accessories according to the information outlined in the chapter "Operation with AC/DC
power adapter";
• Press the switch to position I to turn the device on (the external power pack does not need to be connected)
• The fully charged battery life is about 70 minutes with continuous operation.
WARNING: Before using the device, check the charge level of the lithium-ion battery. Charge the battery
before each use. To keep the device in good working order, charge the battery every 3 months (if not used).
ENGLISH
"Proximity" switch
Adjustment button
ON/OFF switch
depression "+"

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

28183

Inhaltsverzeichnis