Dräger 6510 Gebrauchsanweisung Seite 39

Atemalkohol-messgerät breath alcohol monitor
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Grenze/Beispiel
Limit/Example
PF
Anzeigemodus
PAF
Displaymode
ZPWF
Example for factory's setting.
Factory settings
Pass:
< 0.10
mg/L
Alert:
< 0.24
mg/L
Fail:
< 0.48
mg/L
High:
≥ 0.48
mg/L
Blow criteria
Setting of the minimal breathing volume
and the minimum blowing time for a valid
sample. After selection of the function, en-
try in litres (L) by pressing
.
OK
Factory setting is 1.2 L and 2 seconds.
Control Mode
Increased resolution with 3 decimal places only for measuring in
mg/L and µg/L.
Parameter settings are reset after switching off the instrument.
It is possible to activate all functions simultaneously.
0
Pass
Pass
Zero
Pass
Blow criteria
Back to menu
Min. volume
: 1.2
Min. blow time
: 2.0
,
and
M
< 0,10 mg/L
< 0,24 mg/L
Fehler/Fail
Alarm/Alert
Fehler/Fail
Warn
Fehler/Fail
< 0,48 mg/L
Hoch/High
Hoch/High
Hoch/High
Alkoholkonzentration
Alcohol concentration
39