Herunterladen Inhalt Inhalt Diese Seite drucken

Providing The Breath Sample - Dräger Alcotest 9510 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Alcotest 9510:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NOTICE
i
i
If no entries are made for 10 minutes, the message
DATA INPUT TIMEOUT is displayed, and the
instrument goes back to READY.
Up to 24 letters can be entered for a name. For names
with more than 24 letters/characters, the 24th digit is
the last character entered. The entry is exited with
Next.
The prompt NAME OPERATOR is displayed.
1. Enter the operator name and confirm with Next.
The prompt SUBJECT'S FAMILY NAME is displayed.
2. Enter the surname of the test subject and confirm with
Next.
The prompt SUBJECT'S FIRST NAME is displayed.
3. Enter the first name of the test subject and confirm with
Next.
The prompt SUBJECT'S DATE OF BIRTH is displayed.
4. Enter the digits for the day (DD), month (MM) and year
(YYYY) of birth and confirm with Done.
NOTICE
i
i
If implausible dates are entered, data acceptance is
refused after selecting the Summary button, and a
fresh date entry is requested.
Up to this point, all previous entries can be recalled on
the screen with Prev. The entries can be changed by
making new entries or confirmed with Done.
A summary of the data entered
is displayed. By double-clicking
on the individual entries these
can be edited again.
5. Confirm the entries for the
test
subject
data
pressing Next. The measuring process starts immediately
after the confirmation. Scrolling back through or modifying
previous entries is no longer possible.
NOTICE
i
i
If an external keyboard is used, the data entry can also
be completed using the [Alt] + [s] key combination.
The measuring process is continued.
PURGING is displayed.
The measuring system is purged with ambient air. The
internal zero-point is then measured, BLANK CHECK is
displayed.
A subsequent testing of the ambient air for alcoholic
vapours with the EC sensor only takes place for the
Alcotest 9510 Standard IR / EC, in which case
ENVIRONMENT CHECK is displayed.
If influencing factors in the ambient air can be ruled out in
this test, the instrument will prompt for a breath sample to
be provided. The time displayed in the bottom section
shows the time at which the provision of the breath sample
must be started.
Dräger Alcotest 9510 Standard IR / EC
by
6. Open the packaging of a new mouthpiece, take the breath
hose out of the storage tray and insert the mouthpiece
without skin contact into the handle at the end of the breath
hose using the packaging.
4.2.2

Providing the breath sample

The instrument users must ensure that the person being tested
has not ingested any substances through the mouth or nose
during the control time (at least 10 minutes) before the delivery
of the breath sample.
When the display shows PLEASE BLOW, the instrument is
ready for a breath sample for 3 minutes. Within this time period
the person being tested has to deliver a breath sample.
1. Ask the test subject to
breathe
in
calmly
expire the breath sample
continuously and evenly
into the mouthpiece until
the bottom bar on the
display is filled at least half.
During the delivery of the
breath sample the progress is indicated by the bar graph
being filled.
The length of the bar to be filled is a reference point for the
volume to be supplied until the minimum volume is
reached.
It is, however, possible to continue the delivery of the
breath sample when the bar graph is already filled
completely. This does not impair the measurement.
At the end of the breath sample
If the minimum breath flow is not met, STOP BLOWING is
displayed.
2. End the process of providing the breath sample
immediately.
Restarting (second breath) must always be avoided or the
measurement will be cancelled.
If the requirements for providing the breath sample are met
with regard to duration, volume and continuity, REMOVE
MOUTHPIECE is displayed.
3. Pull the mouthpiece from the handle at the end of the
breath hose with the aid of the mouthpiece packaging and
dispose of it.
4. Place the breath hose back into the storage tray or on the
side of the instrument.
The measuring system is purged with ambient air again
and another internal calibration takes place.
PURGING and BLANK CHECK are displayed in
succession.
If all framework conditions
of the process were met,
the final result is displayed
in the configured unit on the
display.
NOTICE
i
i
Measurement results below 0.03 mg/L or 0.05 % are
output as 0.00 mg/L or %. For other measurement
units corresponding display limits are valid.
Printing of the protocol is started.
Performing the measurement
and
21

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis