Troubleshooting; Error, Cause, Remedy - Dräger Alcotest 9510 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Alcotest 9510:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

When Lab
report is selected, a special protocol with
enhanced sensor data is provided.
Test subject data are not printed in both cases.
7

Troubleshooting

7.1

Error, cause, remedy

Error
»Display« (printer text)
No function after power-on
»ZERO TEST INCORRECT«
(BLANK CHECK INCORRECT)
»MOUTH ALCOHOL«
(MOUTH ALCOHOL)
»MINIMUM VOLUME NOT ACHIEVED«
(MIN VOLUME NOT ACHIEVED)
»BLOWINGTIME TOO SHORT«
(BLOWINGTIME TOO SHORT)
»BLOWING NOT ALLOWED«
(BLOWING NOT ALLOWED)
»READINESS TO BLOW EXPIRED«
(READINESS TO BLOW EXP.)
»ALCOHOL IN THE ENVIRONMENT«
(ALCOHOL IN ENVIRONMENT)
»CHECK AIRWAY«
»DATA INPUT TIMEOUT«
(DATA INPUT TIMEOUT)
»INTERFERENCE«
(INTERFERENCE)
»CALIBRATION EXPIRED (DISABLED)«
The following errors are indicated on the
screen in English language with an error
code and a text message:
»ERROR [XXX] ..
..MAIN SYSTEM«
..IR SYSTEM«
..EC SYSTEM«
..FLOW SYSTEM«
..PRESSURE SYS«
..HEATER SYSTEM«
Example:
ERROR [118] HEATER SYSTEM
The error code and the text message are
printed on the printer.
Example:
Error [118] HEATER SYSTEM
Dräger Alcotest 9510 Standard IR / EC
i
i
Cause
Fuses defective.
Power supply insufficient.
Zero value calibration faulty.
Residual alcohol in the mouth of the test
subject. The test subject has taken or
brought up alcohol immediately prior to
the test.
The provided volume is smaller than the
required minimum volume.
The time for which the breath sample is
delivered is shorter than the minimum
requested.
Breath sample delivery at a not permitted
time, instrument not ready for
measurement (reblow).
Readiness for measurement of
3 minutes has expired. The breath
sample was not started during this time.
Contamination in the ambient air has
been detected.
The purging volume required for purging
the system has not been achieved.
No data input within 10 minutes despite
prompt.
The two independent measuring
systems have detected an impermissible
influence on the measuring values by
foreign substances in the breath.
Calibration interval expired.
Instrument defective
(XXX = code specifying the error)
The instrument will automatically block
all measuring processes if it detects the
error.
The set temperature for the sensor or
hose is not reached.
NOTICE
The control mode and laboratory protocol settings are
disabled after turning off the instrument.
Remedy
Check the fuses, replace if necessary,
see page 26.
Check power supply.
Ensure clean ambient air free from
alcohol. Ventilate test space.
Remedy the cause. Repeat
measurement after a waiting period.
Observe control time.
The instrument automatically prompts for
a repeat of the breath sample.
The instrument automatically requests a
repeat of the breath sample
Instruct the test subject in the correct
provision of the breath sample. Start new
measurement.
Start new measurement.
Ensure clean ambient air free from
alcohol. Ventilate test space.
Check the breath hose inlet. If without
anomalies, start new measurement.
If the error recurs: inform DrägerService.
Start new measurement.
Enter data when prompted to do so.
Identify the cause of this interference
and rectify if possible.
If necessary, trigger other measures to
secure evidence.
Calibrate the instrument or contact
DrägerService.
Turn off the instrument.
Check whether the error returns after
reactivation.
If yes: inform DrägerService. Have the
instrument serviced.
Ensure that the ambient conditions are
met. Switch off instrument, if required put
on instrument cover and switch on again.
If the error should reoccur upon
switch-on, inform Dräger Safety.
Troubleshooting
25