Performing the measurement

Procedure

Prompt the test person to blow evenly
and continuously into the mouthpiece. A
sufficient breath flow is confirmed by a
continuous audible signal and the green
LED.
During the delivery of the breath sample the
following message appears in the display:
» BLOW «.
If the breath sample is sufficient (minimum
breath volume and the minimum blowing
period were reached), the sampling is
triggered. Then the green LED goes out, the
audible signal stops.
The following message appears in the display:
» WAIT Analyzing «

Test Result

After 3 to 20 seconds (dependent on the
detected concentration) the measurement
result appears in the display.
-
The measurement result is displayed with
the corresponding unit.
-
In the default setting, the unit is set to mg/
L (milligram ethyl alcohol per liter
breathing air) and the solution is set to
two positions after the comma. The
default setting can be changed on the
main menu (see Technical Handbook
Dräger Alcotest 7510).
In addition to the measurement result, an
audible signal which is dependent on the
measurement result is output, for example:
measurement result below the limit value (PASS) = 1 x,
measurement result between the limit values (ALERT) = 2 x,
measurement result above the limit value (FAIL) = 3 x
40
If the measurement results are positive, it is recommended to perform
a second measurement after 10 to 15 minutes in order to confirm the
11:53:47
measurement result.
04.12.2017
Documentation of the Results:
A result protocol can be generated with the Dräger Mobile Printer, see
page 42.
BLOW

Additional Measurement

Pressing the
During the recovery time of the sensor (see "Technical Data" on
page 51) the following appears in the display: » WAIT «
i
i
01123996 01 en

Error during the delivery of the breath sample

If the volume of the breath sample is
insufficient, the following appears in the
11:53:47
display:
04.12.17
» Insufficient volume «.
0.00
A short audible signal beeps and the red LED
flashes.
To repeat, press the
mg/L
Waiting period for a new testing standby is 4
to 20 seconds.
In the case of highly irregular exhalation, e.g.
Confirm
sudden termination or suction at the end of
the breath sample, the following appears in
01223996_01_en
the display: » Blow interruption «.
A short audible signal beeps and the red LED flashes.
To repeat, press the
key prepares a new measurement.
OK
NOTICE
Periodically switching off the device does not reduce this
waiting period!
key.
OK
key.
OK
11:53:47
04.12.17
Insufficient
volume
Confirm
01323996_01_en
Dräger Alcotest 7510