5. Transport, packaging and storage

5.2 Packaging and storage

Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during trans-
port (e.g. change in installation site, sending for repair).
Permissible storage temperature
-40 ... +70 °C

6. Commissioning, operation

WARNING!
Physical injuries and damage to property and the
environment caused by media escaping under high
pressure
With the pressurisation of the instrument, as a result of
poor sealing of the process connection, media under high
pressure can escape.
Due to the high energy of the media that can escape in
the event of a failure, the possibility of physical injuries and
damage to property exists.
The sealing of the process connection must be carried out
expertly and checked for leak tightness.

6.1 Mechanical connection

In accordance with the general technical regulations for pressure gauges
(e.g. EN 837-2 "Selection and installation recommendations for pressure
gauges").
Instruments must be grounded via the process connection.
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
EN