Herunterladen Inhalt Inhalt Diese Seite drucken

Bea 060 Configuration; Program Description; Calibration; 17 5.3 Service Intervals - Bosch BEA 060 Originalbetriebsanleitung

Abgasmessgerät für benzinfahrzeuge
Vorschau ausblenden Andere Handbücher für BEA 060:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
4.3.2

BEA 060 configuration

After installing the BEA software, the device interface
of the BEA 060 has to be set. The BEA 060 can be
connected by way of either the USB connecting cable
or Bluetooth to a PC/Laptop.
1. Select "Start >> All programs >> Bosch >>
Diagnostics >> Central Device Communication" or
click onto " " in the task bar with the left mouse
key.
 CDC – Central Device Communication is opened.
2. Select "Settings >> Device interfaces".
i
Pressing <F1> opens the Online Help for a display
of the most important information on interface
setting.
3. Configure the BEA 060 in the GAM Group (USB or
Bluetooth).
 BEA 060 is ready for operation.
4. Start SystemSoft BEA-PC.
4.4

Program description

i
Refer to the Online Help for further details on the
BEA-PC system software.
Robert Bosch GmbH
Operation | BEA 060 | 23 |
5.

Maintenance

!
Work on electrical equipment is only to be
performed by persons with sufficient knowledge
and experience of electrical systems!
5.1
Cleaning
!
Do not use abrasive cleaning agents and coarse
workshop cloths!
The housing and the LCD are only to be cleaned
using a soft cloth and neutral cleaning agents.
5.2

Calibration

For applications demanding mandatory calibration,
regular calibration is essential. (depending on the
regulations of the respective country).
5.3
Service intervals
i
Routine maintenance ensures constant readiness
for operation.
The following maintenance intervals are to be
observed unless otherwise specified by the approval
authorities.
5.3.1
Half-yearly maintenance
Replacement of coarse filter GF1 (fig. 3, pos. 8) in
sampling hose.
Replacement of coarse filter GF2 (fig. 3, pos. 1),
refer to Section 5.4.4.
Check whether both PVC hoses are connected to
the gas outlets (fig. 4, pos. 6 and 7).
Visual inspection of sampling probe (fig. 3, pos. 10).
Sampling probe leak test (refer to Section 5.4.1).
5.3.2
Annual maintenance
This work is to be performed by a qualified
maintenance service. It includes the half-yearly
maintenance and the following additional items:
Checking of BEA 060 measurement accuracy with
a test gas.
Replacement of activated charcoal filter (fig. 3,
pos. 4) in zero gas path.
Replacement of coarse filter GF3 (refer to
Section 5.4.4)
Compliance with the specifications of the relevant
authorities must be ensured.
en
|
1 689 989 182
2020-06-24

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis