Medisana IN 500 Compact Gebrauchsanweisung Seite 29

Vorschau ausblenden Andere Handbücher für IN 500 Compact:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.3
If the dev ce s used 10 t mes a day w th each 20 m nutes of nhalat on
at room temperature (25°C), the expected l fe cycle of the compressor s
L fe cycle
3 years. The accessor es show a d fferent expected l fe cycle:
Nebul zer un t: 100 days
Mouthp ece / Nose adaptor: 3 months
A r tube: 1 year
A r filter: 2 months
Mask: 1 year
4.4
Trouble-
shoot ng
Dev ce s not work ng
after sw tch ng on
No nebul zat on
No nebul zat on or low
nebul zat on rate
No nebul zat on or too
less nebul zat on rate
The dev ce gets very
hot
Problem
The plug s not
correctly connected to
the ma ns outlet.
Vapor zer head s
m ss ng or no resp.
too less med cat on n
the nebul zer tank.
The nebul zer k t s
not correctly
assembled or s t lted
at an ncorrect angle.
The a r tube s not
correctly attached or
s folded or damaged.
The dev ce s covered
or has been used for
more than 20
m nutes.
4 M scellaneous
Poss ble cause
Solut on
Sw tch off the dev ce
("O") and connect the
ma ns plug to a
correspond ng outlet.
Do not use extens on
cords etc. Sw tch on
the dev ce.
Assemble the
nebul zer k t n the
correct way. F ll n
med cat on (m n. 2 ml
/ max. 10 ml) nto the
nebul zer tank.
Follow the nstruct ons
n chapter 3.1
Prepar ng for
nhalat on. Pay
attent on, that the
dev ce s not t lted at
an angle greater than
45° n any d rect on.
Check the correct
connect on of the a r
tube on the
compressor un t and
on the nebul zer k t.
The a r tube must not
be folded, k nked or
bent. Replace the a r
tube f damaged.
Do not cover the
dev ce. Allow the
dev ce to cool down
for at least 40
m nutes.
GB
E
23