Herunterladen Inhalt Inhalt Diese Seite drucken

Eu Declaration Of Conformity; Safety Instructions - IKA RH basic 2 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für RH basic 2:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Contents

EU Declaration of conformity

We declare under our sole responsibility that this product corresponds to the
directives 2014/35/EU, 2006/42/EC, 2014/30/EU and 2011/65/EU and conforms
with the following standards or normative documents: EN 61010-1, EN 61010-2-
010, EN 61010-2-051, EN 61326-1, EN 60529 and EN ISO 12100.
A copy of the complete EU Declaration of Conformity can be requested at
sales@ika.com.
EN

Safety instructions

The IKA RH basic 2; RH basic KT/C or RH digital KT/C was not
Page
constructed for operations in dangerous atmospheres, for mixing
11
dangerous substances or for operation under water.
11
12
› Do not use the device in explosive atmospheres, it is not EX-protected.
12
› With substances capable of forming an explosive mixture, appropriate safety
13
measures must be applied, e.g. working under a fume hood.
13
› To avoid body injury and property damage, observe the relevant safety and
13
accident prevention measures when processing hazardous materials.
14
› Set up the device in a spacious area on an even, stable, clean, non-slip, dry
14
and fireproof surface.
15
Attention - Magnetism! Effects of the magnetic field have to be taken
15
into account (e.g. data storage media, cardiac pacemakers ...).
16
16
› Ensure that the power cord set / temperature sensor cable does not touch the
17
heating plate. The power cable must be replaced by a cable of equal quality
17
and capacity.
17
› Version RH basic KT/C and RH digital KT/C: Examine regularly (we
18
recommend every 4 weeks) the function of the safety temperature limiter. For
this the safety temperature must be changed around at least 1/4 revolution of
the value already stopped and be accomplished the attitude as under "adjusting
the safety temperature delimitation SAFE temp" described. By the exceeding
of the adjusted safety temperature (e.g. rotary button "temp" posed on right
end stop) you can examine the function. The "temp" pilot lamp must expire
and the heating must be lasting switched off.
› When using PTFE-coated magnetic bars, the following has to be noted:
Chemical reactions of PTFE occur in contact with molten or solute alkali metals
and alkaline earth metals, as well as with fine powders of metals in groups 2
and 3 of the periodic system at temperatures above 300 °C - 400 °C. Only
elementary fluorine, chlorotrifluoride and alkali metals attack it; halogenated
hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and "Ulmann", Volume 19)