Assembling the DigiCam Illuminator
4
2
5
1 Plastic cover
2 Connector
3 LED
4 DigiCam Illuminator
5 Centring bolt
6 Slit projector swivel arm
Fig. 19
Assembling the DigiCam Illuminator
Remove the plastic cover (1, Fig. 19) on the swivel arm (6, Fig. 19) of
the slit projector.
Insert the DigiCam Illuminator's (4, Fig. 19) centring bolt (5, Fig. 19)
into the centring hole (7, Fig. 19), such that the LED (3, Fig. 19) faces
towards the patient side.
Insert the DigiCam Illuminator's power cable (2, Fig. 19) into the
relevant socket (10, Fig. 19) on the slit lamp's power supply
(8, Fig. 19).
If the slit lamp does not have a separate power supply (e.g.
examination units), a separate 6 V power supply (11, Fig. 19) is
required for the DigiCam Illuminator. Insert the DigiCam Illuminator
power cable (2, Fig. 19) into the relevant socket (12, Fig. 19) on the
power supply.
000000-1363-329 SL 120 11.07.2006
3
1
6
9
7
7 Centring hole
8 Rear of the power supply
9 Socket for fixation lamp cable
10 Socket for the DigiCam Illuminator
11 Separate power supply (6 V)
12 Connector for the DigiCam Illuminator
Assembly