Herunterladen Diese Seite drucken

Mettler Toledo InPro 4250 Betriebsanleitung Seite 3

Ph-einstabmessketten

Werbung

General instructions for pH combination electrode types InPro
1.
Introduction
These general instructions must be readily available to operators of the electrode at all times. Before using the electrode, these instructions must be read and
understood by all operating personnel. METTLER TOLEDO pH sensors type InPro 4250 and InPro 4250 SG (ISM) are low-maintenance, pressure-resistant
measuring electrodes with a polymer electrolyte XEROLYT
for the simultaneous measurement of pH value and temperature in industrial processes.
The sensor InPro 4250 / InPro 4250 SG (ISM) is suitable for applications involving:
• heavily contaminated media
• emulsions and suspensions
• solutions containing protein
• acid solutions
InPro 4250 / InPro 4250 SG sensors are non-sterilizable /non-autoclavable and unsuitable for measurements, in media containing chlorine or bromine.
2.
Safety remarks
The sensor is intended only for the use described in §1 above. The sensor consists partly of glass. If the sensor is to be cleaned or calibrated using acid or
alkaline solutions, safety goggles and gloves should be worn. In addition, local and in-house-safety regulations must be observed. We recommend that the
electrodes only be put into operation in conjunction with original products/accessories from METTLER TOLEDO. The electrodes may be operated and serviced
only by personnel familiar with the equipment in question and who have read and understood these instructions. Please follow the guidelines enclosed for an
installation in the hazardous areas.
The electrodes are designed to withstand a maximum operational pressure of 8 bar and fulfill all current PED requirements (Pressure Equipment Directive
97/23/EC, Art. 3, Sec. 3.). Please take extra care during installation/fitting of the sensor to ensure that neither the polymer body/shaft of the sensor nor the
thread (Pg 13.5) suffer any injury or damage. Please do not touch or hold the electrode directly at the plug. Electrostatic discharge can damage the chip.
3.
Product description
The markings on each sensor provide the following information:
®
InPro
425X (SG) / XXX / YYY / ZZZ designation (Solution Ground Option) / shaft length in mm / temperature sensor / ISM (option)
combination pH
Pt100 / Pt1000
pH 0 ...14 / pH 1 ...14
0 ...130 °C
II 1/2G EEx ia IIC T6/T5/T4/T3 Ex classification
SNCH 00 ATEX 3130 X
IS/I, II, III/1/ABCDEFG/T6
Follow the operating instructions note
Order No. 52 002 14X
Additionally, the sensor is supplied with a serial number on the connector cap for identification purposes for each sensor produced.
4.
Installation and preparation for use
1.
On unpacking, check the sensor for mechanical damage of the pH-sensitive glass and shaft. Report any damage immediately to your METTLER TOLEDO
supplier.
2.
Remove the watering cap and rinse the sensor with de-ionized water. After rinsing, the sensor should only be dabbed dry with a tissue. Do not rub the
glass membrane, since this can lead to electrostatic charging and sluggish response times.
3.
Check the membrane chamber of the glass electrode for the presence of any air bubbles and remove same by lightly swinging the sensor in a vertical
plane.
4.
Please refer to the instruction manual of the relative housing in order to install the sensor.
5.
Connect the sensor to the pH transmitter, using a cable with the appropriate connector. Please refer to the wiring and terminal diagram accompanying
the cable.
6.
For InPro 4250 SG electrodes, with an additional Pt auxiliary electrode, connect the corresponding blue wire in the VP cable to your transmitter in
accordance with instructions in the connection diagram of VP cable and/or in the transmitter instruction manual. For ORP measurement (sep. transmitter
needed) connect solution ground (blue wire) with ORP input.
7.
InPro 4250 SG electrodes with ISM allow "Plug and Measure" as well as advanced diagnostics. Operation of the electrodes in ISM mode requires the
use of a special VP8 cable as well as an ISM pH-module for the transmitter M 700 (software). For installation, startup and operation of the ISM system
please refer to the relative instructions provided with the cable and the module.
5.
Calibration and measurement
InPro 4250 sensors require 2-point calibration. To calibrate, first remove the watering cap. The sensor is then dipped in succession into two different buffer
solutions with given pH reference values and the transmitter calibrated to these buffer values (for example buffer pH 4.01 an d pH 7.00). Please refer to the
instruction manual of the transmitter for details.
6.
Maintenance
The sensor with polymer electrolyte XEROLYT
to recondition it before use by immersion in Friscolyt™ (Order no. 51 340 053) for at least 24 hours. After each working cycle, the sensor tip and the aper-
tures (diaphragms) must be thoroughly cleaned using de-ionized water.
7.
Trouble-shooting
1.
Sluggish response time.
The glass membrane can be reactivated using a HF solution (order no. 51 319 053). Only that section of the sensor tip below the two (diaphragms)
apertures may be immersed in the HF solution, and this during a period of about 5 minutes. Immediately afterwards, rinse the sensor thoroughly and
immerse in Friscolyt™ for at least 12 hours.
2.
Insufficient sensitivity (slope).
Check the apertures for contamination and rinse with warm water. Check the VP connector for traces of moisture. If necessary, clean well using de-
ionized water or alcohol, and afterwards dry carefully.
8.
Disposal
If no relative official regulations are in force, used or defective InPro 4250 / InPro 4250 SG sensors and the packaging can be disposed of as ordinary house-
hold waste.
9.
Warranty
In the event of manufacturing faults, 12 months warranty from date of production is granted.
®
PLUS and a built-in temperature sensor. InPro 4250 / InPro 4250 SG (ISM) sensors are designed
• sulfide-bearing media
• hot alkaline (caustic) solutions
• processes with wide pressure fluctuations
• solvents
type of sensor
type of built-in temperature sensor
pH measuring range InPro 4250 / InPro 4251 and InPro 4252 (SG) (with reduced performance
and lifetime at pH 12 ... 14)
temperature range for operation (reduced performance and lifetime at temperatures > 110 °C [InPro 4250],
resp. > 80 °C [InPro 4251/InPro 4252])
designation and number of certificate
FM approved
ordering number
®
PLUS cannot be refilled. The sensor must not be stored dry. In the event that the sensor dries out, it is necessary
®
®
4250 and InPro
4250 SG (ISM)

Werbung

loading

Diese Anleitung auch für:

Inpro 4250 sgInpro 4250 sg ism