8. Rotate the left knob (C) to the "BACK" icon.
9. Briefly press the left knob (C) to return to the top
level of the settings menu.
10. Rotate the left knob (C) to scroll to 'EXIT'.
11. Briefly press the left knob (C) to return to the main
operating screen.
Note: There is not an icon on the display to
indicate that Power Recovery has been activated.

2.10 Reset to Factory Default Settings

Resetting the unit to Factory Default Settings will do the following:
Clear all single point calibration (SPC) temperatures.
Turn off Power Recovery.
Re-enable the beeper setting.
1. To reset the unit to factory default settings, the unit
must first be in the top level of the settings menu.
2. Rotate the left knob (C) to scroll to the 'SYS'
(System) setting.
3. Briefly press the left knob (C) to enter the System
settings menu.
4. Rotate the left knob (C) to the "RSET" (Reset)
setting.
EN-29