SETTING

4.1 Access to mode "Operating Data"

When using the centrifuge, the following parameters can be set:
While the centrifuge is turned off, press simultaneously the keys "Time" (7) and "Stop/Lid" (9) and turn on the
main switch of the centrifuge. Now release both keys and as a result a display test is executed for approx. 5
seconds. All indicators will appear at the same time (See figure 27).
ATTENTION:
After the settings have been stored, change back to normal mode again by switching off the centrifuge
for a short while!
M11
-
Number of Starts
-
Operating Hours of Centrifuge
-
Operating Hours of Motor
-
Software Version
-
Error List
-
Function of the Imbalance Switch
-
Operation of Keyboard
-
Hardware Version
-
Intermediate Circuit Voltage in Volt