Herunterladen Inhalt Inhalt Diese Seite drucken

Medel ELITE Handbuch Seite 13

Vorschau ausblenden Andere Handbücher für ELITE:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
4.4 How to use the memory function:
• In Clock mode ( Off mode) , press MEM button. The number of stored data
is showed (pic14), sudden the latest result will be displayed with date and
time (pic15).
• Press MEM button repeatedly to review the results measured .
• How to see the average of the last 3 measurements done with T3 System.
(The average is calculated upon last 3 readings measured within 15 minutes).
Please take the measurements following the instructions in the SECTION T3
System of this manual; then, in power off mode, press T3 Button and the
average value of the last 3 measurements made within 15 min. with T3
System, will be displayed with AVG symbol blinking (pic. 16).
NOTE:
The average data using T3 button is shown only if the last 3 measurements have
been done within 15 minutes. Otherwise, the data will be shown as 0.
4.5 To delete all values stored in the memory:
In Clock mode ( Off mode) press MEM button , after the latest data stored will be
displayed press again and hold MEM button for 3 sec. All results will be deleted
display. Press the button MEM or O/I, the display will turn off. (pic.17)
5. ERROR MESSAGE/MALFUNCTION
Follow the below recommended actions and press O/I button again to initiate
another measurement.
Error code
Er 0
Pressure system is unstable
before measurement
Er 1
Fail to detect systolic pressure
Er 2
Fail to detect diastolic pressure
Er 3
Pneumatic system blocked or
cuff is too tight during inflation
Possible cause(s)
Recommended Action
Don't move and try again.
Apply the cuff correctly and try
again
EN
13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis