Power
30 A/m
frequency (50/
60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Guidance and manufacturer's declaration – electromagnetic immunity
The UPA-WU10T/UPA-WU10R is intended for use in the electromagnetic environment
specified below. The customer or the user of the UPA-WU10T/UPA-WU10R should assure that
it is used in such an environment.
IEC 60601
Immunity test
test level
Conducted RF
3 Vrms
150 kHz to 80 MHz
IEC 61000-4-6
outside ISM bands
6 Vrms
150 kHz to 80 MHz
in ISM bands
30 A/m
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the UPA-WU10T/UPA-WU10R,
including cables, than the recommended
separation distance calculated from the
equation appliance to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 √P
3 Vrms
c
6 Vrms
c
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.