Instructions for Use
Installation
Install this device indoors only, and in any case do not use the pump
in an environment exposed to atmospheric agents (rain, ice, snow),
dust, aggressive gases, or in explosive environments or those with a
high fire risk.
During operation, the following environmental conditions must be
respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The TwisTorr 704 FS pump must be used in conjunction with one of
the suitable Agilent controller and must be connected to a primary
pump (see "Technical Information").
The available controllers are the following:
Tab. 1
CONTROLLER
Controller Rack
Controller onboard
The TwisTorr 704 FS complies to installation category II as per directive EN
CAUTION!
61010- 1. Connect the device to a mains line that satisfy the above category.
For installation of optional accessories, see "Technical Information".
264/334
maximum pressure: 2 bar above atmospheric pressure;
temperature: from +5 °C to +35 °C;
relative humidity: 0 – 90 % (non-condensing).
PART NUMBER
X3501-64016
X3512-64016
TwisTorr 704 FS User Manual / 87-901-051-01 (A)
PRODUCT NAME
TwisTorr Medium-TMP
Rack Controller
TwisTorr Medium TMP
onboard Controller