EN-2

2. INSTALLATION

2.1 Select the Location
Avoid heat sources, rapid temperature changes, air current or excessive vibrations. Allow sufficient
space.
2.2 Leveling
Be sure the balance is level before it is used or after its location
is changed.
The PJX balance has a level bubble in a small round window
beside the display.
To level the balance, adjust the 4 leveling feet until the bubble is
centered in the circle.
Please refer to the right figure for leveling.
2.3 Connecting Power and Acclimatising the Balance
Connect the DC output connector to the power receptacle on the rear of the balance. Then connect the
AC adapter plug to a suitable electrical outlet.
Acclimatising
It is suggested that the balance should not be used until it has been connected to power and
acclimatised to the environment for a certain period of time. In the case of a balance with the precision
above 0.1 mg, the acclimatisation time should be 1.5 hours; in the case of balance with the precision of
0.01 mg, the acclimatisation time should be more than 4 hours.
2.4 Connecting the Interface
The PJX balance has two connectivities, RS232 port and Mini DIN.
Use the RS-232 port to connect either to a computer or a printer with a standard (straight-through) serial
cable.
Mini DIN is used to connect with another accessory, as an alternative option for RS232.
Interface connections on the rear of the balance.
Mini DIN
RS232
RS232: Used to connect to PC or Printer
Note: See the Printing section for Connecting, Configuring and Testing the Printer / Computer Interface.